James Long Company
Lab Tips
-
James Long Company distributes updates in the ".ZIP"
compressed form. Windows has built-in support for decompressing ".ZIP"
files.
-
The magnitude of 60 Hz (or other environmental noise) present at the
bioamplifier output is influenced by the following considerations.
-
First the magnitude is proportional to the 60 Hz (or other noise)
field strength present both "common mode" and "differential mode".
Second, fields may be either induced into loops or dipoles, the
effects of each type of field is different as are the remedies.
Common mode v. differential mode and loop v. dipole are discussed
below.
-
Common mode voltage is that induced voltage present at both the plus
and minus inputs of a channel referenced to ground. Common mode 60 Hz
is present because all instrumentation (differential) amplifiers
require a ground reference connection for the amplifiers to work
correctly. (This is because instrumentation amplifiers can tolerate
common mode voltages of only up to the internal power supply voltage,
usually a few volts; without a ground
connection the common mode voltage might rise to tens or hundreds
of volts.) A mathematically ideal amplifier would not pass any
common mode voltages because the amplifier is supposed to amplify
only differences between the plus and minus input; the perfect amplifier
should not pass voltages that are differences between both
inputs and ground. Real amplifiers do pass common mode voltages.
Common mode 60 Hz at the inputs can be minimized by dressing the
ground lead close by the other electrode leads and choosing a
ground site that is nearby all the other sites, e.g., AFz. (If only
ECG were being recorded then the best ground site would be midway
between the two active electrodes.) The effects of 60 Hz (or other
frequency) common mode noise vary greatly between amplifiers.
The deleterious effects of common mode noise are specified by the
amplifier manufacturer as a logarithmic ratio. Our bioamps have a
Common Mode Rejection Ratio (CMRR) of roughly 112 dB measured at 60
Hz. Older amplifiers have a CMRR in the 80 dB to 90 dB range;
this means the effect on spectral power of common mode noise is 150
to 1500 times worse on an older bioamplifier; hence, the required
but distorting 60 Hz notch filter. Note: some amplifiers
specifications report CMRR at alpha frequencies; this greatly
overstates CMRR because CMRR deteriorates with increasing
frequency. CMRR also deteriorates with greater impedance mismatch
between the plus and minus inputs, a specification rarely reported by
bioamplifier manufacturers.
-
Differential mode 60 Hz is that induced voltage that presents a
potential difference between the plus and minus inputs of a channel.
Amplifier design can not change this voltage. This is the very
voltage the amplifier is supposed to amplify.
-
Other factors that also directly affect 60 Hz noise magnitude are:
-
Electrode lead dressing and induced loop voltages.
-
Including all low-level signals in one ribbon cable greatly
reduces the area of any electrical loops in space. This
argues for use of electrode drops from the cap for facial EMG
and EOG signals.
-
Further, the effective area of any loop can be pushed towards
zero by adding a few twists to the ribbon cable as it runs from
the subject to the headbox and from the headbox to the bioamp.
Each twist results in induced voltages
180 degrees out of phase with the adjacent twist,
so induced signals tend to cancel out.
-
Note that induced loop voltages do not directly depend on
electrode impedance.
-
Electrode impedance
-
Impedance mismatch between the plus and minus input of a channel (as
stated above) diminishes CMRR. So, while an amplifier may have
a published CMRR of 112 dB, the actual CMRR may be far below
that if the impedance mismatch is large enough. In labs with
high noise levels this argues for neither collecting referenced
to A1/A2 nor for re-referencing to A1/A2 since it is harder to
achieve as low impedance in A1 and A2 as scalp sites.
-
Dipole voltages do not depend on closing of a loop. They can
appear as a voltage difference between two wires unconnected to
each other. Since the ideal bioamplifier has infinite input
impedance, the bioamplifier would see the full amount of this
induced voltage. In practice, the two wires would be connected
to electrodes attached to the subject. Thus, most induced voltage would
be shunted through the subject instead of appearing at the
bioamplifier inputs. The degree to which any induced voltage is
shunted through the subject is inversely proportional to
electrode impedance. This alone is the major reason for
attempting to reduce electrode impedance. Recall that induced
loop voltages will not be affected by electrode impedance.
Indeed, if a small coil, e.g., a transformer, were near the
electrode leads inducing a voltage into the lead loop, no
amount of electrode impedance reduction would reduce the
induced voltage.
-
Electrode impedance used to be an important factor in obtaining
accurate voltage measurements. Ancient amplifiers
had input impedances of one megaohm or
ten megaohm. Our amplifier has an input impedance well in
excess of one gigaohm. In the old days, if one electrode,
say F3 had an impedance of 1 kiloohm and the homologous
electrode, F4, had an impedance of 11 kiloohms and the bioamp
had an input impedance of 1 megaohm then the bioamp's load
on the scalp signal would attenuate the F4 site by 1 percent
voltage (2 percent in power) relative to the F3 site. This
would tend to confound any hemispheric laterality study
(indeed it could confound any study). If the bioamplifier
input impedance is very large relative to the electrode
impedance, e.g., one gigaohm v. ten kiloohm (or even one
hundred kiloohm) then electrode impedance no longer impacts
measured voltage. Such is the case with our amplifiers.
-
Bioamplifier intrinsic noise.
All bioamplifiers add a noise voltage to the signal. This noise
is primarily thermal (3/2 kT noise) and, thus, is spectrally
white or flat. This noise is usually specified by the
bioamplifier manufacturer. The amplitude units would be in
volts per radical hertz.
(Less commonly, the noise would be
specified in power units, e.g., square volts per hertz or
watt ohms per hertz.)
Care must be taken to ensure amplitude specifications are comparable.
First, specifications must be adjusted so that the measurement
bands are comparable. Second, amplitudes must be adjusted if
one specification is in peak-to-peak amplitude and another is
in rms (root mean square) amplitude.
Amplifiers have other noise sources,
e.g., 1/f noise; this noise tends only to dominate
at very low frequencies and, thus, only affects protocols such
as CNV studies.
-
There is no amplitude measurement reason for
matching impedances between electrodes.
Also, electrically linking
A1 with A2 has largely been abandoned;
that practice required nearly perfect
impedance matching. Instead, researchers are deriving A1+A2/2
algebraically, e.g., with EEGTRANS. Finally, if the reason for
matching impedance is in some hope of matching induced noise
levels, this hope is for naught because, as stated above, the
induced noise voltage has very much to do with lead dressing and
orientation in space, so the homologous sites will never match in
noise. Thus, a better rule is: if the noise power is
essentially insignificant in the band of interest (either directly
or via aliasing), it does not matter what the impedance differences
are; the goal should be to establish impedance limits that ensure
insignificant noise levels.
Copyright (c) 2006 James McMartin Long.